Abstract

We describe an all−in−fibre apparatus for Constant Intensity Direct Absorption Spectroscopy (CIDAS) for gas concentration measurements which keeps the power of a diode laser constant along the frequency sweep. The reduction of the large variation of the laser power, connected to the frequency scan, enhances the ability of detecting small variations in a background signal, resulting in an increase of the sensitivity with respect to standard direct absorption techniques. Moreover, CIDAS allows for a real−time observation of the absorption signals without any kind of post−detection processing. The apparatus has been tested with carbon dioxide (CO2) and methane (CH4), around 1.57 and 1.65 μm, respectively.

Go to article

Authors and Affiliations

A. Montori
M. Pas de
M. Giuntini
M. Siciliani Cumis de
S. Viciani
F. D’Amato