Abstract

The following paper presents an idea of minimising the number of connections of individual piezoelectric transducers in a row-column multielement passive matrix system used for imaging of biological media structure by means of ultrasonic projection. It allows to achieve significant directivity with acceptable input impedance decrease. This concept was verified by designing a model of a passive ultrasonic matrix consisting of 16 elementary piezoceramic transducers, with electrode attachments optimised by means of electronic switches in rows and columns. Distributions of acoustic field generated by the constructed matrix model in water and results of the calculations conformed well.

Go to article

Authors and Affiliations

Krzysztof Opieliński
Tadeusz Gudra
Piotr Pruchnicki
Download PDF Download RIS Download Bibtex

Abstract

Mixed boundary-value problem for periodic baffles in acoustic medium is solved with help of the method developed earlier in electrostatics. The nice feature of the method is that the resulting matrices are relatively easy for computations and that the results satisfy exactly the energy conservation law. Illustrative numerical examples present the wave-beam steering (in the far-field) in a baffle system that may be considered as a model of one-dimensional ultrasonic transducer array.

Go to article

Authors and Affiliations

Yuriy Tasinkevych
Eugeniusz Danicki