Abstract

Synthetic aperture (SA) technique is a novel approach to present day commercial systems and has previously not been used in medical ultrasound imaging. The basic idea of SA is to combine information acquired simultaneously from all directions over a number of emissions and to reconstruct the full image from these data.

The paper presents the multi-element STA (MSTA) method for medical ultrasound imaging. The main difference with the STA approach is the use of a few elements in the transmit mode in contrast to a single element aperture. This allows increasing the system frame rate, decreasing the number of emissions, and provides the best compromise between the penetration depth and lateral resolution. Besides, a modified MSTA is proposed with a corresponding RF signal correction in the receive mode, which accounts for the element directivity property.

In the experiments a 32-element linear transducer array with 0.48 mm inter-element spacing and a burst pulse of 100 ns duration were used. Two elements wide transmission aperture was used to generate an ultrasound wave covering the full image region. The comparison of 2D ultrasound images of a tissue mimicking phantom obtained using the STA and MSTA methods is presented to demonstrate the benefits of the second one.

Go to article

Authors and Affiliations

Ihor Trots
Marcin Lewandowski
Yuriy Tasinkevych
Andrzej Nowicki
Download PDF Download RIS Download Bibtex

Abstract

The following paper presents an idea of minimising the number of connections of individual piezoelectric transducers in a row-column multielement passive matrix system used for imaging of biological media structure by means of ultrasonic projection. It allows to achieve significant directivity with acceptable input impedance decrease. This concept was verified by designing a model of a passive ultrasonic matrix consisting of 16 elementary piezoceramic transducers, with electrode attachments optimised by means of electronic switches in rows and columns. Distributions of acoustic field generated by the constructed matrix model in water and results of the calculations conformed well.

Go to article

Authors and Affiliations

Krzysztof Opieliński
Tadeusz Gudra
Piotr Pruchnicki
Download PDF Download RIS Download Bibtex

Abstract

The paper presents the theoretical and experimental study of synthetic transmit aperture (STA) method combined with Golay coded transmission for medical ultrasound imaging applications. The transmission of long waveforms characterized by a particular autocorrelation function allows to increase the total energy of the transmitted signal without increasing the peak pressure. It can also improve signal-to-noise ratio and increase the visualization depth maintaining the ultrasound image resolution.

In the work the 128-element linear transducer array with 0.3 mm pitch excited by the 8 and 16-bits Golay coded sequences as well as a one cycle at nominal frequencies 4 MHz were used. The comparison of 2D ultrasound images of the tissue mimicking phantoms is presented to demonstrate the benefits of coded transmission. The image reconstruction was performed using synthetic STA algorithm with transmit and receive signals correction based on a single element directivity function.

Go to article

Authors and Affiliations

Ihor Trots
Andrzej Nowicki
Yuriy Tasinkevych
Marcin Lewandowski
Download PDF Download RIS Download Bibtex

Abstract

Mixed boundary-value problem for periodic baffles in acoustic medium is solved with help of the method developed earlier in electrostatics. The nice feature of the method is that the resulting matrices are relatively easy for computations and that the results satisfy exactly the energy conservation law. Illustrative numerical examples present the wave-beam steering (in the far-field) in a baffle system that may be considered as a model of one-dimensional ultrasonic transducer array.

Go to article

Authors and Affiliations

Yuriy Tasinkevych
Eugeniusz Danicki
Download PDF Download RIS Download Bibtex

Abstract

The paper presents the optimization problem for the multi-element synthetic transmit aperture method (MSTA) in ultrasound imaging applications. The optimal choice of the transmit aperture size is made as a trade-off between the lateral resolution, penetration depth and the frame rate. Results of the analysis obtained by a developed optimization algorithm are presented. The maximum penetration depth and lateral resolution at given depths are chosen as optimization criteria. The results of numerical experiments carried out in MATLAB® using synthetic aperture data of point reflectors obtained by the FIELD II simulation program are presented. The visualization of experimental synthetic aperture data of a tissue mimicking phantom and in vitro measurements of the beef liver performed using the SonixTOUCH Research system are also shown.

Go to article

Authors and Affiliations

Marcin Lewandowski
Yuriy Tasinkevych
Andrzej Nowicki
Ihor Trots