Evaluation of portable blood glucose meters using canine and feline pooled blood samples

Journal title

Polish Journal of Veterinary Sciences

Yearbook

2016

Issue

No 4

Authors

Divisions of PAS

Nauki Biologiczne i Rolnicze

Publisher

Polish Academy of Sciences Committee of Veterinary Sciences ; University of Warmia and Mazury in Olsztyn

Date

2016

Identifier

DOI: 10.1515/pjvs-2016-0089 ; ISSN 1505-1773

Source

Polish Journal of Veterinary Sciences; 2016; No 4

References

Domori (2014), The clinical utility of two human portable blood glucose meters in canine and feline practice, Vet Clin Pathol, 43, 55, doi.org/10.1111/vcp.12115 ; Wess (2000), Capillary blood sampling from the ear of dogs and cats and use of portable meters to measure glucose concentration, J Small Anim Pract, 41, 60, doi.org/10.1111/j.1748-5827.2000.tb03164.x ; Freckmann (2012), System accuracy evaluation of blood glucose monitoring systems for self - monitoring of blood glucose according to DIN EN ISO, Diabetes Sci Technol, 6, 15197. ; Mori (2011), Evaluation of artificial pancreas technology for continuous blood glucose monitoring in dogs, J Artif Organs, 14, 133, doi.org/10.1007/s10047-011-0560-1 ; Wess (2000), a Assessment of five portable blood glucose meters for use in cats, Am J Vet Res, 61, 1587, doi.org/10.2460/ajvr.2000.61.1587 ; Barman (1993), Separation and size distribution of red blood cells of diverse size shape , and origin by flow / hyperlayer field - flow fractionation, Anal Biochem, 212. ; Brito (2014), ISO - based assessment of accuracy and precision of glucose meters in dogs, J Vet Intern Med, 28, 1405, doi.org/10.1111/jvim.12397 ; Higgins (1982), Glycosylated hemoglobin in human and animal red cells Role of glucose permeability, Diabetes, 31, 743, doi.org/10.2337/diab.31.9.743 ; Clarke (1987), Evaluating clinical accuracy of systems for self - monitoring of blood glucose, Diabetes Care, 10, 622, doi.org/10.2337/diacare.10.5.622 ; Cox (1997), Understanding error grid analysis, Diabetes Care, 20, 911, doi.org/10.2337/diacare.20.6.911 ; Johnson (2009), Comparison of a human portable blood glucose meter , veterinary portable blood glucose meter , and automated chemistry analyzer for measurement of blood glucose concentrations in dogs, J Am Vet Med Assoc, 235. ; Khositseth (2015), Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion, J Clin Monit Comput, 29, 145, doi.org/10.1007/s10877-014-9579-1 ; Rendell (1985), An interspecies comparison of normal levels of glycosylated hemoglobin and glycosylated albumin, Comp Biochem Physiol, 819. ; Magnani (1978), Comparative studies on red blood cell glucose phosphorylating activities of mammals, Comp Biochem Physiol, 323.
×